Programming Guide for DM3058/DM3058E
3-27
1. :MEASure?
Command Format:
:MEASure?
Function:
The command queries if a new data has been acquired under current trigger
setting.
Return Format:
The query returns TRUE or FALSE. Thereinto, TRUE indicates a new data has
been acquired.
2. :MEASure
Command Format:
:MEASure {AUTO|MANU}
Function:
The command selects the measurement type as Auto or Manual.
Example:
Sets the measurement type to Manual.
:MEASure MANU
3. :MEASure:VOLTage:DC?
Command Format:
:MEASure:VOLTage:DC?
Function:
The command queries the current measured DC voltage value.
Explanations:
The command is unavailable in double-display.
Return Format:
Seitenansicht 51