
RIGOL Chapter 2 Performance Calibration Test
DS1000E/D Performance Verification Guide
2-8
23) Output a sine waveform with 1 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
24) Repeat step 6).
25) Output a sine waveform with 20 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
26) Repeat step 8), 9) and 10).
27) Output a sine waveform with 50 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
28) Repeat step 12), 13) and 14).
29) Turn CH1 off. Test CH2 according to the above test steps and record the test results.
Kommentare zu diesen Handbüchern