RIGOL
Users Guide for DS1000B Series
2-51
Key Points:
Real-time Sampling:
DS1000B has Real-time sampling rate up to 2GSa/s. At the time base 20ns
or faster, the oscilloscopes use the sine(x)/x interpolation to expand the
horizontal time base.
Equivalent sampling:
Known as Repetitive sampling to get up to 20ps of horizontal resolution
(equivalent 50Gsa/s). This mode is good for observing repetitive signals,
and it is not recommended for single-shot or pulse.
Normal Acquisition:
Oscilloscope acquires signal by equal time interval.
Average Acquisition:
Apply averaging to your signal to remove uncorrelated noise and improve
measurement accuracy. Reduces random or uncorrelated noise in the signal
display. The averaged waveform is a running average over a specified
number of acquisitions from 2 to 256.
Peak Detect:
Peak Detect mode captures the maximum and minimum values of a signal,
and finds highest and lowest record points over many acquisitions.