Rigol DM3062 Betriebsanweisung Seite 70

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 138
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 69
RIGOL
©Copyright RIGOL Technologies, Inc. 2007.
User’s Guide for DM3000 Series
2-20
Auto Triggering
Auto triggering takes continuous readings at the fastest rate possible for the
specified measurement configuration (function, range, resolution, and so forth).
Auto trigger is a default trigger mode when the multimeter power-on.
Press Æ Auto, enter the menu shown below:
Figure 2-19
Table 2-16 Auto Trigger Function Menu Explanation
Interval time:
You can manually specify a delay between the trigger signal and the first sample
that follows. This may be useful in applications where you want to allow the input
signal to settle before taking a reading, or for pacing a burst of readings.
• The trigger delay may be set from 400 to 2000 ms.
• The continuity and diode test functions ignore the trigger delay segment.
• If a trigger delay is not manually set, the default trigger delay is automatically set.
• If you manually specify a trigger delay, that delay is used for all measurement
functions (except continuity and diode test).
Function
Menu
Setting Explanation
Interval
Set interval time in 400~2000ms.
Hold
Save all changes, back to a higher level menu.
Seitenansicht 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 137 138

Kommentare zu diesen Handbüchern

Keine Kommentare