RIGOL User’s Guide DS1000B Series Digital Oscilloscopes DS1074B,DS1104B,DS1204B Dec. 2014 RIGOL Technologies, Inc.
RIGOL User’s Guide for DS1000B Series VIII Contents Guaranty and Declaration ...
RIGOL User’s Guide for DS1000B series 2-70 2. Remote Control (1) By Web Page When the network configuration is completed, open the IE and enter t
RIGOL User’s Guide for DS1000B Series 2-71 (2) By Commands DS1000B is a USB TMC (USB Test & Measurement Class) instrument, which could be contro
RIGOL User’s Guide for DS1000B series 2-72 Figure 2-93 VISA TCP/IP Resource interface Click “Next” and select “Auto-detect of LAN Instrument”.
RIGOL User’s Guide for DS1000B Series 2-73 The software searching the available network device automatically. Figure 2-95 Search the network dev
RIGOL User’s Guide for DS1000B series 2-74 Figure 2-97 Communication testing dialog Make a connection by USB Connect the oscilloscope to a
RIGOL User’s Guide for DS1000B Series 2-75 Figure 2-98 Measurement & Automation Explorer interface Select the responding resource and open t
RIGOL User’s Guide for DS1000B series 2-76 Language The DS1000B series oscilloscopes have multi-language user menu, choose as your desire. Press Ut
RIGOL User’s Guide for DS1000B Series 2-77 Pass/Fail The Pass/Fail function monitors changes of signals by comparing the input signal which is within
RIGOL User’s Guide for DS1000B series 2-78 1. Mask setting Press Utility→Pass/Fail→Mask Setting to go to the following menu. Figure 2-103 Table
RIGOL User’s Guide for DS1000B Series 2-79 Figure 2-105 Table 2-59 The Mask setting menu (External, Page 2/2) Menu Settings Comments Save Go to
RIGOL User’s Guide for DS1000B Series IX Waveform Recorder ... 2-80 Print Setting .
RIGOL User’s Guide for DS1000B series 2-80 Waveform Recorder Waveform recorder records input waveforms from CH1, CH2, CH3 and CH4 with a maximum rec
RIGOL User’s Guide for DS1000B Series 2-81 Figure 2-108 Table 2-62 The Record menu (Page 2/2) Menu Settings Comments Interval <1.00ms-1000s>
RIGOL User’s Guide for DS1000B series 2-82 Run/Stop Button The Run/Stop button can also replay or continue the waveform display. The Current Frame w
RIGOL User’s Guide for DS1000B Series 2-83 Figure 2-113 Table 2-67 The Storage menu (External location Page 2/2) Menu Settings Comments Save Go
RIGOL User’s Guide for DS1000B series 2-84 Print Setting DS1000B series oscilloscopes support two print modes: “normal” and “PictBridge”. Please exe
RIGOL User’s Guide for DS1000B Series 2-85 2. To set up print parameters: Press Utility→Print set to into a setting menu. Please setup the printer
RIGOL User’s Guide for DS1000B series 2-86 Figure 2-118 Table 2-70 Print setup menu (PictBridge mode, Page 2/4) Menu Settings Comments Paper si
RIGOL User’s Guide for DS1000B Series 2-87 3. Print: Make sure the Printer has been connected and the print settings are all ready before printing
RIGOL User’s Guide for DS1000B series 2-88 Self-Calibration The Self-Calibration adjusts the internal circuitry to get the best accuracy. Use this f
RIGOL User’s Guide for DS1000B Series 2-89 System Information Press Utility→System Info to view the system information, including Model, Serial N
RIGOL User’s Guide for DS1000B series 2-90 Preference Press Utility→Preference to go to the following menu Figure 2-123 Table 2-73 The Preference
RIGOL User’s Guide for DS1000B Series 2-91 Date and Time DS1000B Series provide build-in system real-time date and time. Users can set the date and t
RIGOL User’s Guide for DS1000B series 2-92 To Measure Automatically The Measure button in the menu area activates the automatic measurement functi
RIGOL User’s Guide for DS1000B Series 2-93 Figure 2-128 Table 2-76 The Measure menu (Page 2/2) Menu Settings Comments Display All OFF ON Turn off al
RIGOL User’s Guide for DS1000B series 2-94 Quick Measurement Setup Press Measure→QuickMea button, go to the quick measure menu。 Figure 2-129 Table
RIGOL User’s Guide for DS1000B Series 2-95 Delay/Phase Setup Press Measure→Delay/Phase Setup button, and go to the delay/phase setup menu. Figure 2-
RIGOL User’s Guide for DS1000B series 2-96 Voltage Measurements The DS1000B series oscilloscopes provide automatic voltage measurements including Vp
RIGOL User’s Guide for DS1000B Series 2-97 Press Measure→Voltage button, and go to the following menu. Figure 2-132 Table 2-79 The Voltage Measurem
RIGOL User’s Guide for DS1000B series 2-98 Time Measurements The DS1000B series oscilloscopes provide 12 kinds of timing parameters auto-measure
RIGOL User’s Guide for DS1000B Series 2-99 Press Measure→Time button, and go to the following menu Figure 2-136 Table 2-82 The Time Measurement me
RIGOL User’s Guide for DS1000B Series 1-1 Chapter 1 Quick Guide This chapter covers the following topics: General Inspect Panels and User I
RIGOL User’s Guide for DS1000B series 2-100 Key Points: 1. Select the signal channel for measuring. CH1, CH2, CH3, CH4 according to the interest.
RIGOL User’s Guide for DS1000B Series 2-101 To Measure with Cursors Figure 2-140 shows the Cursor button on the front-panel. Figure 2-140 The Cur
RIGOL User’s Guide for DS1000B series 2-102 Manual Mode Press Cursor→Mode→Manual to display the manual mode menu. Figure 2-141 Table 2-85 The Curs
RIGOL User’s Guide for DS1000B Series 2-103 Table 2-86 The Cursor menu Cursor Increment Operation Cursor A X Y Turn the multifunctional knob () to
RIGOL User’s Guide for DS1000B series 2-104 Track Mode Press Cursor→Mode→Track to display the manual mode menu. Figure 2-142 Table 2-87 Cursor des
RIGOL User’s Guide for DS1000B Series 2-105 Table 2-88 The Cursor usage Cursor Operation Cursor A Turn the multifunctional knob ( ) to move cursor
RIGOL User’s Guide for DS1000B series 2-106 Auto mode Press Cursor→Mode→Auto to display the manual mode menu. Figure 2-143 Table 2-89 The Mode set
RIGOL User’s Guide for DS1000B Series 2-107 To Use Run Control Buttons The RUN control buttons include Auto (auto setting), Run/Stop and Single.
RIGOL User’s Guide for DS1000B series 2-108 Auto-Set functions When the Auto is pressed the oscilloscope is configured to the following defaults: Ta
RIGOL User’s Guide for DS1000B Series 3-1 Chapter 3 Application Examples Example 1: Taking Simple Measurements The function is used to observe an
RIGOL User’s Guide for DS1000B series 1-2 General Inspect After receiving a new DS1000B series oscilloscope, please inspect the instrument as foll
RIGOL User’s Guide for DS1000B Series 3-2 Example 2: View a Signal Delay Caused by a Circuit Th
RIGOL User’s Guide for DS1000B Series 3-3 Example 3: Capture a Single-Shot Signal To capture a single event, it needs to gather some pre-test knowl
RIGOL User’s Guide for DS1000B Series 3-4 Example 4: To Reduce the Random Noise on a Signal If
RIGOL User’s Guide for DS1000B Series 3-5 4. To reduce the noise by setting the acquisition type and adjust the waveform intensity. (1) If there is
RIGOL User’s Guide for DS1000B Series 3-6 Example 5: Making Cursor Measurements There are 22 b
RIGOL User’s Guide for DS1000B Series 3-7 Measure the Amplitude of the First Waveform Peak of the Sinc. Please follow these steps: 1. Press Cursor k
RIGOL User’s Guide for DS1000B Series 3-8 Example 6: The Application of the X-Y operation View
RIGOL User’s Guide for DS1000B Series 3-9 Figure 3-6 Ellipse method to observe the phase difference Sinθ = A/B or C/D, where θ = phase shift (in de
RIGOL User’s Guide for DS1000B Series 3-10 Example 7: Triggering on a Video Signal Test a vide
RIGOL User’s Guide for DS1000B Series 3-11 Triggering on the Video Lines 1. Press the MENU key in the Trigger control area to see the trigger menu.
RIGOL User’s Guide for DS1000B Series 1-3 Panels and User Interface Being familiar with the operation panels is the first priority when you get a
RIGOL User’s Guide for DS1000B Series 3-12 Example 8: FFT Cursor Measurement FFT measurements
RIGOL User’s Guide for DS1000B Series 3-13 Example 9: Pass/Fail Test In Pass/Fail Test, the oscilloscope compares the input signal to the establis
RIGOL User’s Guide for DS1000B Series 4-1 Chapter 4 Troubleshooting 1. After the oscilloscopes is powered on, th
RIGOL User’s Guide for DS1000B Series 4-2 the Trigger Mode as “AUTO”. Moreover, push Auto button
RIGOL User’s Guide for DS1000B Series 5-1 Chapter 5 Specifications All specifications apply to the DS1000B Series
RIGOL User’s Guide for DS1000B Series 5-2 Technical Specifications Acquisition Sampling Modes R
RIGOL User’s Guide for DS1000B Series 5-3 Sample Rate and Delay Time Accuracy ±50ppm (over any≥1ms time interval) Ver
RIGOL User’s Guide for DS1000B Series 5-4 Trigger Trigger Sensitivity 0.1div-1.0div (adjustable
RIGOL User’s Guide for DS1000B Series 5-5 Reciprocal of ∆T in Hertz (1/∆T) Track Voltage value for Y-axis waveform Ti
RIGOL User’s Guide for DS1000B series 1-4 Front Panel Interfaces: USB Host: Connect DS1000B with external USB devices when the oscilloscope is regar
RIGOL User’s Guide for DS1000B Series 5-6 General Specifications Display Display Type 5.7 inch.
RIGOL User’s Guide for DS1000B Series 6-1 Chap
RIGOL User’s Guide for DS1000B Series 6-2 Appendix B: Warranty RIGOL warrants its products’ mainframe and accessories in materials and workmanship
RIGOL User’s Guide for DS1000B Series 6-3 Appe
RIGOL User’s Guide for DS1000B Series 1 Index
RIGOL . User’s Guide for DS1000B Series 2 UTILITY ... 2-64 Vertical ... 5-3 Vertic
RIGOL User’s Guide for DS1000B Series 1-5 Notation definitions in this Manual: Throughout this guide, notation symbols of buttons and knobs are the
RIGOL User’s Guide for DS1000B series 1-6 User Interface Figure 1-3 User Inte
RIGOL User’s Guide for DS1000B Series 1-7 Functional Check Perform this quick functional check to verify that the instrument is operating correct
RIGOL User’s Guide for DS1000B series 1-8 Figure 1-5 Probe connection Figure 1-6 Probe adjustment [1] Attach the probe tip to the Pr
RIGOL User’s Guide for DS1000B Series 1-9 To Compensate the Probes Perform this adjustment to match the characteristics of the probe and the channe
RIGOL User’s Guide for DS1000B series 1-10 To Display a Signal Automatically The oscilloscope has an automated feature to display the input signal
RIGOL User’s Guide for DS1000B Series 1-11 To Understand the Vertical System As shown in Figure 1-8, there are several buttons and knobs in the VE
RIGOL User’s Guide for DS1000B series 1-12 Measurement hints If the channel is DC coupled, measuring the DC components of the signal by simply notin
RIGOL User’s Guide for DS1000B Series 1-13 To Understand the Horizontal System Figure1-9 shows the HORIZONTAL controls: MENU button, and kn
RIGOL User’s Guide for DS1000B series 1-14 3. The horizontal knob moves displayed signal horizontally on the waveform window. The horizontal knob
RIGOL User’s Guide for DS1000B Series 1-15 To Understand the Trigger System Figure 1-10 shows the TRIGGER control: MODE, MENU, FORCE, 50%
RIGOL User’s Guide for DS1000B series 1-16 3. Press MENU button in the Trigger control area. A soft button menu appears on the display showing the
RIGOL User’s Guide for DS1000B Series 1-17 To Understand the Quick Function The following figure shows that there are two buttons in the QUICK func
RIGOL User’s Guide for DS1000B Series I Guaranty and Declaration Copyright © 2009 RIGOL Technologies, Inc. All Rights Reserved. Trademark Inform
RIGOL User’s Guide for DS1000B Series 2-1 Chapter 2 Operating Your Oscilloscope By now, a user should understand the VERTICAL, HORIZONTAL and TRI
RIGOL User’s Guide for DS1000B series 2-2 To Set up the Vertical System Four analog channels are provided by DS1000B: CH1, CH2, CH3 and CH4. Pres
RIGOL User’s Guide for DS1000B Series 2-3 Knobs Explanation Use the vertical controls to display signal waveforms by adjusting the vertical knob, a
RIGOL User’s Guide for DS1000B series 2-4 Channel Settings Each channel of DS1000B series digital oscilloscope has an independent operation menu an
RIGOL User’s Guide for DS1000B Series 2-5 Figure 2-2 Table 2-3 Channel menu (Page 2/2) Menu Settings Comments 2/2 Back to the previous menu page
RIGOL User’s Guide for DS1000B series 2-6 1. To set up channel coupling To use Channel 1 as an example, input a square wave signal with DC shift.
RIGOL User’s Guide for DS1000B Series 2-7 Press CH1→Coupling→GND, to set “GND” coupling, it disconnects the input signal. The screen displays as Figu
RIGOL User’s Guide for DS1000B series 2-8 2. To set up the channel bandwidth limit Take Channel 1 for an example, input a signal containing high f
RIGOL User’s Guide for DS1000B Series 2-9 3. To set up Probe Attenuation The oscilloscope allows selecting the attenuation factor for the probe. Th
RIGOL User’s Guide for DS1000B Series II Safety Notices view the following safety precautio
RIGOL User’s Guide for DS1000B series 2-10 4. Digital Filter Press CH1→ Digital Filter to display the digital filter menu. Figure 2-9 Turn off d
RIGOL User’s Guide for DS1000B Series 2-11 Figure 2-11 Table 2-5 The Filter menu Menu Settings Comments Digital Filter ON OFF Turn on the digital
RIGOL User’s Guide for DS1000B series 2-12 5. To change the Volts/Div settings The Volts/Div control has Coarse or Fine configuration. The Vertica
RIGOL User’s Guide for DS1000B Series 2-13 6. To invert a waveform Invert turns the displayed waveform 180 degrees with respect to the ground level
RIGOL User’s Guide for DS1000B series 2-14 Math Function The mathematic functions include “add”, “subtract”, “Multiply” and “FFT” for CH1, CH2, CH3
RIGOL User’s Guide for DS1000B Series 2-15 Figure 2-17 Table 2-7 The Math menu (Page 2/2) Menu Settings Comments The multifunctional knob ( ) ad
RIGOL User’s Guide for DS1000B series 2-16 1. Using FFT function The FFT (Fast Fourier Transform) process converts a time-domain signal into its f
RIGOL User’s Guide for DS1000B Series 2-17 Key points for FFT 1. Signals that have a DC component or offset can cause incorrect FFT waveform compone
RIGOL User’s Guide for DS1000B series 2-18 2. Selecting an FFT Window DS1000B series oscilloscope provides four FFT windows. Each window is a tra
RIGOL User’s Guide for DS1000B Series 2-19 Term Explanations FFT Resolution: The FFT resolution is the quotient between sampling rate and the number
RIGOL User’s Guide for DS1000B Series III interior of the device or electric shock, please do not operate in a humid environment. Do Not Operate in
RIGOL User’s Guide for DS1000B series 2-20 REF Function Reference Waveforms are saved waveforms to be selected for display. The reference function w
RIGOL User’s Guide for DS1000B Series 2-21 Figure 2-21 Table 2-12 REF menu when using internal/external memory (Page 2/3) Menu Settings Comments S
RIGOL User’s Guide for DS1000B series 2-22 1. Import and Export Switch to external memory and select Internal location. Then, press REF →Imp. /Exp
RIGOL User’s Guide for DS1000B Series 2-23 2. Export Switch to external memory and select Internal location. Then, press REF→Imp. /Exp.→ Export
RIGOL User’s Guide for DS1000B series 2-24 3. Save to External Memory Switch to external memory and select External location. Then, press REF→Save
RIGOL User’s Guide for DS1000B Series 2-25 4. New File (or New Folder) Switch to external memory and select External location. Then, press REF→S
RIGOL User’s Guide for DS1000B series 2-26 5. Import Switch to external memory and select External location. Then, press REF→Import and enter t
RIGOL User’s Guide for DS1000B Series 2-27 6. Display a Reference Waveform Figure 2-34 Reference waveform display Operation introductions: 1.
RIGOL User’s Guide for DS1000B series 2-28 To Set up the Horizontal System Press MENU button to enter the horizontal setting system. Figure 2-3
RIGOL User’s Guide for DS1000B Series 2-29 Term Explanations Slow Scan Mode: This mode is available when the horizontal time base is set to 50ms or sl
RIGOL User’s Guide for DS1000B Series IV Safety Terms and Symbols Terms in this Manual. Thes
RIGOL User’s Guide for DS1000B series 2-30 Knobs Explanation The horizontal controls change the horizontal scale and position of waveforms. The hori
RIGOL User’s Guide for DS1000B Series 2-31 Delayed Scan The Delayed Scan is a magnified portion of the main waveform window. Use Delayed Scan to loca
RIGOL User’s Guide for DS1000B series 2-32 X-Y Format This format is useful for studying phase relationships between CH1 and ch2 (CH3 and CH4 can’t
RIGOL User’s Guide for DS1000B Series 2-33 To Set up the Trigger System The trigger determines when the oscilloscope starts to acquire data and dis
RIGOL User’s Guide for DS1000B series 2-34 Edge Trigger An edge trigger determines whether the oscilloscope finds the trigger point on the rising or
RIGOL User’s Guide for DS1000B Series 2-35 Pulse Width Trigger Pulse trigger occurs according to the width of pulse. The abnormal signals can be dete
RIGOL User’s Guide for DS1000B series 2-36 Video Trigger Choose video trigger to trigger on fields or lines of NTSC, PAL, or SECAM standard video si
RIGOL User’s Guide for DS1000B Series 2-37 Figure 2-43 Table 2-25 The Video Trigger menu (Line Num, Page 2/2) Menu Settings Comments Line Num <
RIGOL User’s Guide for DS1000B series 2-38 Select “Line Synchronization”: Select “Filed Synchronization”: Figure 2-45
RIGOL User’s Guide for DS1000B Series 2-39 Pattern Trigger Pattern trigger recognises the trigger condition through inspecting the code preset which
RIGOL User’s Guide for DS1000B Series V General-Purpose Oscilloscopes The introduction of this document contains three types of DS1000B series osci
RIGOL User’s Guide for DS1000B series 2-40 Figure 2-48 Pattern trigger: set code as falling edge
RIGOL User’s Guide for DS1000B Series 2-41 Alternative Trigger When alternative trigger is on, the trigger sources come from two vertical channels. T
RIGOL User’s Guide for DS1000B series 2-42 Figure 2-51 Table 2-30 The Alternative trigger menu (Page 1/2 Trigger Type: Pulse) Menu Settings Comme
RIGOL User’s Guide for DS1000B Series 2-43 Figure 2-53 Table 2-32 The Alternative trigger menu (Page 1/2 Trigger Type: Video) Menu Settings Commen
RIGOL User’s Guide for DS1000B series 2-44 Trigger Settings Set up different trigger settings according to different trigger modes. In Edge, Pulse
RIGOL User’s Guide for DS1000B Series 2-45 Figure 2-56 Table 2-35 Page 2/2 Settings for Edge, Pulse and Pattern Triggers Menu Settings comments Ho
RIGOL User’s Guide for DS1000B series 2-46 Trigger Holdoff You can use trigger Holdoff to stabilize a complex waveform, such as a pulse sequence. Ho
RIGOL User’s Guide for DS1000B Series 2-47 Trigger Terms Explanation 1. Trigger Source: Trigger occurs from several sources: Input channels (CH1, CH
RIGOL User’s Guide for DS1000B series 2-48 display becomes stable on the screen. Any factor results in the instability of waveforms can be detected b
RIGOL User’s Guide for DS1000B Series 2-49 5. Adjustable trigger sensitivity To avoid the influence of noise from the physical world, and get the st
RIGOL User’s Guide for DS1000B Series VI Automatic cursor tracking measurements. Wavefor
RIGOL User’s Guide for DS1000B series 2-50 To Set up the Sampling System Figure 2-59 shows, the Acquire button at the MENU of the front panel.
RIGOL User’s Guide for DS1000B Series 2-51 The waveform displayed on the screen will change in conjunction with the setting of Acquire menu. To re
RIGOL User’s Guide for DS1000B series 2-52 Figure 2-63 Adopt Normal acquisition Figure 2-64 Adopt peak detect acquisition, signal contains envel
RIGOL User’s Guide for DS1000B Series 2-53 Key Points: Real-time Sampling: DS1000B has Real-time sampling rate up to 2GSa/s. At the time base 20ns o
RIGOL User’s Guide for DS1000B series 2-54 To Set up the Display System The menu button for the display system on the front panel is shown in the f
RIGOL User’s Guide for DS1000B Series 2-55 Figure 2-67 Table 2-39 The Display menu (Page 2/3) Menu Settings Comments Grading ON OFF The waveform
RIGOL User’s Guide for DS1000B series 2-56 Term Explanations Refresh rate: It is an important performance of digital oscilloscopes. It means the
RIGOL User’s Guide for DS1000B Series 2-57 To Store and Recall Figure 2-69 shows the menu button for the storage system on the front panel. Figure
RIGOL User’s Guide for DS1000B series 2-58 Figure 2-71 Table 2-42 The Storage menu (Bitmap storage) Menu Settings Comments Storage 8-Bitmap 24-Bit
RIGOL User’s Guide for DS1000B Series 2-59 Figure 2-73 Table 2-44 The Storage menu (CSV storage) Menu Settings Comments Storage CSV Create or de
RIGOL User’s Guide for DS1000B Series VII Document Overview Chapter 1 Quick Start To introduce the panls and menus to help users to realize the b
RIGOL User’s Guide for DS1000B series 2-60 Internal Memory Press Storage→Internal to go to the following menu. Figure 2-75 Table 2-46 The Memory
RIGOL User’s Guide for DS1000B Series 2-61 External Memory Press Storage→ External to go to the following interface. Figure 2-76 External memory in
RIGOL User’s Guide for DS1000B series 2-62 U-disk Selection Figure 2-78 U-disk selection DS1000B series have two USB Host ports on the front and r
RIGOL User’s Guide for DS1000B Series 2-63 Disk Management Press Storage→Disk Mgr.. Then go to the following menu. Figure 2-79 Table 2-48 The Disk
RIGOL User’s Guide for DS1000B series 2-64 1. Disk Information Press Storage →Disk Mgr. →Disk info to go to the disk information interface. Fi
RIGOL User’s Guide for DS1000B Series 2-65 Figure 2-83 Table 2-50 The Rename menu Menu Settings Comments To move the input focus point of file n
RIGOL User’s Guide for DS1000B series 2-66 To Set up the Utility Figure 2-69 shows the menu button for the utility system on the front panel. Fi
RIGOL User’s Guide for DS1000B Series 2-67 Figure 2-86 Table 2-52 The Utility menu (Page 2/3) Menu Settings Comments Pass/Fail Setup Pass/Fail
RIGOL User’s Guide for DS1000B series 2-68 The I/O Setup Press Utility→IO Setting to go to the following menu. Figure 2-88 Table 2-54 The I/O Set
RIGOL User’s Guide for DS1000B Series 2-69 1. LAN Setting Access the oscilloscope into Network, and press LAN Set to enter the setting interface, s