Rigol M300 Bedienungsanleitung Seite 25

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 144
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 24
RIGOL
© 2007 RIGOL Technologies, Inc.
User’s Guide for DM3000 Series
1-15
To Measure Resistance
The following shows the system connections and selection of measurement functions.
The practice provides a guide get familiar with the Resistance measurement
technique. Resistance measurement methods include 2-Wire Resistance
Measurement and 4-Wire Resistance Measurement, and will explain
separately.
2-Wire Resistance Measurement
Figure 1- 18 2-Wire Resistance Measurement Interface
Table1- 5 Resistance Measurement Characteristics
Seven Ranges
200Ω, 2kΩ, 20kΩ, 200kΩ, 1MΩ, 10MΩ, 100MΩ
Max Resolution
100uΩ
Open-circuit
Voltage
<7V
Input Protection
1000V on all ranges (HI Terminal)
Configurable
Parameters
Range, Null value
Steps:
1. Connect test leads as shown in Figure 1- 19; red test lead to the HI Terminal,
black test lead to the LO Terminal.
2. Press to select the 2-Wire Resistance Measurement.
3. Choose the appropriate measurement range.
4. Set the Null value
Seitenansicht 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 143 144

Kommentare zu diesen Handbüchern

Keine Kommentare