User’s Guide for DM3000 Series
1-27
Period Test
Figure 1- 34 Period Measurement Data Interface
Table1- 11 Period Test Characteristics
Range
200mV, 2V, 20V, 200V, 750V
Measurement Range
0.33s ~ 3.3us
Input Signal Range
100mVAC~750VAC
Import Protection
750VRMS on all ranges (HI Terminal)
Configurable Parameters
Null value
Steps:
1. Connect test leads as Figure 1- 35 shown. Red test lead connects the HI
Terminal, Black test lead connects the LO Terminal.
2. Press twice to select the Period Test.
3. Set the Null value.
Null computing will be an optional operation, could be setup in accordance with
user demand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set up Measurement Parameters”, Null computing)