RIGOL
Users Guide for DM3058
2-16
4-Wire Resistance Measurement
When the measured resistance is smaller than 100, if both the resistance of Test
leads and the contact resistance between probe and testing point are too big to be
ignored as comparing with measured resistance, the error will be increased
provided that you insist on measuring by via 2-Wire Resistance. So please change it
to 4-Wire.
Operating Steps:
1. Connecting the red leads (Test leads) to Terminal Input-HI and HI Sense and
black leads to Terminal Input-LO and LO Sense as shown in Figure2-20. That is
connecting the Test leads to both side of the measured resistance.
Figure 2-20 Sketch map for connecting
2. Press twice to switch to 4-Wire Resistance Mode.
Figure 2-21 Measurement Interface
3. Choose an appropriate resistance range according to its scope.
I
Resistance
4-Wire Sense LO
4-Wire Sense HI
Seitenansicht 35