RIGOL
Programming Guide for DM3058
5-49
[SENSe:]CURRent:AC:NULL[:STATe]{ON|OFF}
[SENSe:]CURRent:AC:NULL[:STATe]?
Function:
The command sets or queries the enable state of ACI NULL function.
Explanations:
[SENSe:]CURRent:AC:NULL[:STATe]{ON|OFF}: Enable or disable ACI NULL
function.
[SENSe:]CURRent:AC:NULL[:STATe]?: Query the enable state of ACI NULL
function.
The command [SENSe:]CURRent:AC:NULL:VALue can be used to set NULL
value.
The commands CONFigure:CURRent:AC and MEASure:CURRent:AC? can
disable NULL function (OFF).
Return Format:
The query returns the enable state of ACI NULL function.
3. [SENSe:]CURRent:AC:NULL:VALue
Command Format:
[SENSe:]CURRent:AC:NULL:VALue{<value>|MIN|MAX}
[SENSe:]CURRent:AC:NULL:VALue?[{MIN|MAX}]
Function:
The command sets or queries the NULL value of ACI measurement.
Parameters:
Name
Type
Range of Values
Default
<value>
Discrete
± 120% max range
0
Explanations:
[SENSe:]CURRent:AC:NULL:VALue{<value>|MIN|MAX}: Set the NULL value of
ACI measurement. MIN is -120% max range of current function, MAX is 120%
max range of current function, DEF = 0.
[SENSe:]CURRent:AC:NULL:VALue?[{MIN|MAX}]: Query the NULL value of
Seitenansicht 164