Rigol DM3058 Spezifikationen Seite 216

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 264
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 215
RIGOL
Programming Guide for DM3058
5-100
1. TRIGger:COUNt
Command Format:
TRIGger:COUNt{<count>|MIN|MAX|DEF|INFinity}
TRIGger:COUNt?[{MIN|MAX}]
Function:
The command sets or queries the trigger times used in multimeter.
Parameters:
Name
Type
Range of Values
Default
<count>
Consecutive
Integer
1~50000
Required
parameter
Explanations:
TRIGger:COUNt{<count>|MIN|MAX|DEF}: Set the trigger times.
TRIGger:COUNt?[{MIN|MAX}]: Query the trigger times.
The command is only available for remote operation, not for front panel
operation.
The command ―CONFigure‖ or ―MEASure?‖ can be used to set trigger times as 1.
It is suggested that the ―SENSe:<function>:RANGe:AUTOOFF‖ should be used
to disable autorange function, the command ―SENSe:<function>:RANGe‖ ,
―CONFigure‖ or ―MEASure?‖ be used to set range by manual.
Return Format:
The query returns the trigger times used in multimeter.
2. TRIGger:DELay
Command Format:
TRIGger:DELay{<seconds>|MIN|MAX|DEF}
TRIGger:DELay?[{MIN|MAX}]
Function:
The command sets or queries the delay time between trigger signal and starting
measurement.
Seitenansicht 215
1 2 ... 211 212 213 214 215 216 217 218 219 220 221 ... 263 264

Kommentare zu diesen Handbüchern

Keine Kommentare