Programming Guide for DM3058
3-49
9. :trigger:vmcomplete:pulsewidth
Command Format:
:trigger:vmcomplete:pulsewidth?
:trigger:vmcomplete:pulsewidth<value>
Function:
The command queries and sets VMC output pulsewidth.
Parameters:
Name
Type
Range
Default
<value>
consecutive integer
See Explanations
100
Explanations:
:trigger:vmcomplete:pulsewidth?: Query the VMC output pulsewidth. The default
is 100 ms.
:trigger:vmcomplete:pulsewidth<value>: Set the VMC output pulsewidth. The
range of output pulsewidth is vary with sample rate. When the sample is S, the
range is 1ms~400ms; when the sample is M, the range is 1ms~200ms; when the
sample is F, the range is 1ms~30ms.
Return Format:
The query returns the current VMC output pulsewidth.