Rigol DM3068 Betriebsanweisung Seite 158

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 171
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 157
RIGOL Chapter 6 Specifications
User’s Guide for DM3068
6-12
[3]
Relative to calibration standards.
Measuring Characteristics
Frequency and Period
Measurement Method Reciprocal-counting technique, AC-coupled input using the AC voltage function.
Input Impedance 1 ± 2%, in parallel with <150 pF capacitance on any range
Input Protection 750 V rms on all ranges
Measurement Considerations
All frequency counters are susceptible to error when measuring lowvoltage, lowfrequency signals. Shielding inputs from external noise pickup is critical for
minimizing measurement errors.
Settling Considerations
Errors will occur when attempting to measure the frequency or period of an input following a dc offset voltage change. The input blocking RC time constant
must be allowed to fully settle (about 1 s) before the most accurate measurements are possible.
Seitenansicht 157
1 2 ... 153 154 155 156 157 158 159 160 161 162 163 ... 170 171

Kommentare zu diesen Handbüchern

Keine Kommentare