RIGOL
© Copyright RIGOL Technologies, Inc. 2007. 2-99
User’s Guide for DS1000 Series
Menus and operations of Cursor Measurements
1. Manual Mode
Figure 2-120 Table 2-79
Menu Settings Comments
Mode Manual
Adjust the cursor to measure X/Y
parameters manually
Type
X
Y
Shown as vertical line to measure the
horizontal parameters
Shown as horizontal line to measure
the vertical parameters
Source
CH1
CH2
MATH
FFT
LA
Select the measurement signal source.
(LA is only applicable in Mixed Signal
Oscilloscope)
In this mode, the oscilloscope measures the cursors’ coordinate values for Y or X,
and the increment between the two cursors.
To do manual Cursor Measurements, please do as the following steps:
1. Select the Manual mode for cursor measurement by pressing soft button as:
Cursor→Mode→Manual.
2. Select channel Source for measurements by pressing soft button as:
Cursor→Source→CH1, CH2, MATH (FFT), or LA (Mixed Signal Oscilloscope)
NOTE: While measuring the channel of MATH, the results are valued with “d”
(division) as units.
3. Select the cursors type by pressing soft button as Cursor→Type→X or Y.
4. Move the cursors to adjust the increment between the cursors: (Details in the
following Table)
Kommentare zu diesen Handbüchern