Rigol DS1000M Betriebsanweisung Seite 149

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 168
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 148
RIGOL
© Copyright RIGOL Technologies, Inc. 2007. 3-13
User’s Guide for DS1000 Series
Example 9: Pass/Fail Test
The Pass/Fail Test is one of enhanced special functions based on DS5000 series.
When running test function the oscilloscope automatically examines the input signal,
compared to the already built waveform mask. If the waveform “touches” the mask,
a “Fail” occurs, otherwise the test passes. When needed, a programmable output can
be used for external automatic control applications, such as the automatic failure
product kick-out in a production line. The output is built in as a standard feature, and
is optically isolated.
Do the steps as following:
1. Press UtilityPass/Fail
2. Press Enable Test and select ON to open the test.
3. Press Mask SettingLoad
4. Press Load to recall the saved mask or press X Mask and Y Mask to adjust the
horizontal limit and vertical limit then press Create Mask to create a new mask.
4. Press Output to select the expected outputting waveforms.
5. Press Operate to start the test.
Figure 3-11
Seitenansicht 148
1 2 ... 144 145 146 147 148 149 150 151 152 153 154 ... 167 168

Kommentare zu diesen Handbüchern

Keine Kommentare