Rigol DS1000Z Series Spezifikationen Seite 17

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 172
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 16
Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-3
:SINGle
Syntax
:SINGle
Description Set the oscilloscope to single trigger mode. This command is equivalent to the following
two operations: pressing the SINGLE key at the front panel, sending
the :TRIGger:SWEep SINGle command.
Explanation In single trigger mode, the oscilloscope triggers once the trigger conditions are met
and then stops.
When the waveform record function is enabled or during playback the recorded
waveform, this command is invalid.
Related
Command
:TFORce
:RUN
:STOP
:TFORce
Syntax
:TFORce
Description Generate a trigger signal forcefully. This command is only applicable to normal and single
trigger mode (see the
:TRIGger:SWEep command) and is equivalent to press the FORCE
key at the front panel.
Related
Command
:TRIGger:SWEep
Seitenansicht 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 171 172

Kommentare zu diesen Handbüchern

Keine Kommentare