RIGOL Chapter 2 Command System
2-4 DS1000Z Programming Guide
:ACQuire Commands
:ACQuire commands are used to set or query the memory depth, acquisition mode and the current sample
rate of the oscilloscope as well as to set the number of averages under average acquisition mode.
Command List
[1]
:
:ACQuire:AVERages
:ACQuire:MDEPth
:ACQuire:TYPE
:ACQuire:SRATe?
Note
[1]
: In the Command Listin this manual, the parameters in the setting commands and the query
commands are not included and you can refer to the complete introductions of the commands in the text
according to the keyword.
:ACQuire:AVERages
Syntax :ACQuire:AVERages <count>
:ACQuire:AVERages?
Description
Set or query the number of averages under average acquisition mode.
Parameter
Name
Type
Range
Default
<count>
Integer
2
n
(n is an integer and ranges from 1 to 10)
2
Explanation You can sent the :ACQuire:TYPE command to set the acquisition mode.
In average acquisition mode, the greater the number of averages is, the lower the
noise will be and the higher the vertical resolution will be but the slower the
response of the displayed waveform to the waveform changes will be.
Return
Format
The query returns an integer between 2 and 1024.
Example
:ACQuire:AVERages 128 /*Set the number of averages to 128*/
:ACQuire:AVERages? /*The query returns 128*/
Related
Command
:ACQuire:TYPE