Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-57
:MEASure:SETup:MIN
Syntax :MEASure:SETup:MIN <value>
:MEASure:SETup:MIN?
Description Set or query the lower limit of the threshold when measuring time, delay and phase
parameters. The default unit is %.
Parameter
Name
Type
Range
Default
<value>
Integer
5 to 93
10
Explanation The upper limit, middle point and lower limit of the threshold are used to define the
time, delay and phase parameters. Setting these values will affect the measurement
results of the time, delay and phase parameters.
Setting the lower limit higher than the middle point will automatically increase the
middle point to keep it igher than the upper limit.
Return
Format
The query returns an integer between 5 and 93.
Example
:MEASure:SETup:MIN 93
Related
Command
:MEASure:SETup:MAX
:MEASure:SETup:MID
:MEASure:ITEM
:MEASure:SETup:PSA
Syntax :MEASure:SETup:PSA <source>
:MEASure:SETup:PSA?
Description
Set or query the source A of Phase 12
and Phase 12 measurements.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Return
Format
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example
:MEASure:SETup:PSA CHANnel1
Related
Command
:MEASure:ITEM
Seitenansicht 70