22. :MEASure:PPHAse?
Command Format:
:MEASure:PPHAse? [<source A>,<source B>]
Function:
The command measures the phase difference between <sourceA> and
<sourceB> relative to the rising edge. The <source> may be CHANnel1,
CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as <-1.000 e-004, and the unit is s.
23. :MEASure:NPHAse?
Command Format:
:MEASure:NPHAse? [<source A>,<source B>]
Function:
The command measures the phase difference between <sourceA> and
<sourceB> relative to the falling edge. The <source> may be CHANnel1,
CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as <-1.000 e-004, and the unit is s.
24. :MEASure:TOTal
Command Format:
:MEASure:TOTal {{1|ON}|{0|OFF}}
:MEASure:TOTal?
Function:
The command sets all the measurement functions on or off.
Returned Format:
The query returns 1 or 0, respectively indicates ON or OFF.