Rigol DS1204B Spezifikationen Seite 90

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 152
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 89
RIGOL Command Systems
Programming Guide for DS1000B Series
2-78
contains two waveform points
(2) Time(ind) = TimeSpan + TimeOffset
Example: When sample rate is 250k Sa/s and trigger offset is 500 us.
To calculate time of index point 2:
TimeSpan = (2 4096)/(250*2) = -8.188ms
Time(2) = (-8.188+0.5) = -7.688ms
To calculate time of index point 3:
TimeSpan =((3 4096)/2)/(250) = (-2047)/250 = -7.686ms
Time(3)=(-8.186+0.5) = -7.686ms
Time(3) Time(2).
5. Data displays on the screen:
6. Data format of returned screen data:
The screen data obtained by command :WAV:DATA? in Peak Detect mode are stored in
the form of T1max T1min,T2max-T2min, the format of which is the same as memory
data storage. Because one time contains two points, the waveform data extends to 1200
600*2.