RIGOL Chapter 2 Command System
Programming Guide for DM3000
Syntax
:MEASure?
Function Query whether the current measurement is finished. If yes, the query
returns TRUE; otherwise, the query returns FALSE.
2.
:MEASure
Syntax
:MEASure {AUTO|MANU}
Function Set the measurement mode to Auto or Manual.
Default AUTO
3.
:MEASure:VOLTage:DC?
Syntax
:MEASure:VOLTage:DC?
Function The query returns the measurement value of DC voltage in scientific
notation (for example, +2.53021747E-04) and the unit is V.
4.
:MEASure:VOLTage:DC
Syntax
:MEASure:VOLTage:DC {0|1|2|3|4|MIN|MAX|DEF}
Function
Set the range of DC voltage measurement.
Explanation The measurement mode will change to “Manual” while you set the
range.
The range corresponding to each parameter is as follows
Parameter Range
DM306x DM305x
0 200 mV 400 mV
1 2 V 4 V
2 20 V 40 V
3 200 V 400 V
4 1000 V 1000 V
MIN 200 mV 400 mV
MAX 1000 V 1000 V
DEF 20 V 40 V
Example Set the range to the minimum:
:MEASure:VOLTage:DC 0 or
:MEASure:VOLTage:DC MIN
5. :MEASure:VOLTage:DC:RANGe?
Syntax
:MEASure:VOLTage:DC:RANGe?