Rigol M300 Bedienungsanleitung Seite 20

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 89
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 19
RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-8
1. :MEASure?
Syntax
:MEASure?
Function Query whether the current measurement is finished. If yes, the query
returns TRUE; otherwise, the query returns FALSE.
2.
:MEASure
Syntax
:MEASure {AUTO|MANU}
Function Set the measurement mode to Auto or Manual.
Default AUTO
3.
:MEASure:VOLTage:DC?
Syntax
:MEASure:VOLTage:DC?
Function The query returns the measurement value of DC voltage in scientific
notation (for example, +2.53021747E-04) and the unit is V.
4.
:MEASure:VOLTage:DC
Syntax
:MEASure:VOLTage:DC {0|1|2|3|4|MIN|MAX|DEF}
Function
Set the range of DC voltage measurement.
Explanation The measurement mode will change to Manualwhile you set the
range.
The range corresponding to each parameter is as follows
.
Parameter Range
DM306x DM305x
0 200 mV 400 mV
1 2 V 4 V
2 20 V 40 V
3 200 V 400 V
4 1000 V 1000 V
MIN 200 mV 400 mV
MAX 1000 V 1000 V
DEF 20 V 40 V
Example Set the range to the minimum:
:MEASure:VOLTage:DC 0 or
:MEASure:VOLTage:DC MIN
5. :MEASure:VOLTage:DC:RANGe?
Syntax
:MEASure:VOLTage:DC:RANGe?
Seitenansicht 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 88 89

Kommentare zu diesen Handbüchern

Keine Kommentare