RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-40
1. :CALCulate:FUNCtion
Syntax :CALCulate:FUNCtion?
:CALCulate:FUNCtion
{NONE|NULL|DB|DBM|MIN|MAX|AVERAGE|TOTAL|LIMIT}
Function Select the math operation function.
The query returns the current operation function, for example, NULL.
Explanation The definition of each parameter is as follows. For math operations
not supported by the current measurement function, the multimeter
does not support the corresponding commands.
Value Explanation Remark
NONE Disable the math operation
function
Supported by all the
functions
NULL NULL operation Not supported by
continuity and diode
measurement functions
DB dB operation Only supported by DCV
and ACV
DBM dBm operation Only supported by DCV
and ACV
MIN minimum operation Not supported by
continuity and diode
measurement functions
MAX maximum operation Not supported by
continuity and diode
measurement functions
AVERAGE average operation Not supported by
continuity and diode
measurement functions
TOTAL total operation Not supported by
continuity and diode
measurement functions
LIMIT limit operation Not supported by
continuity and diode
measurement functions
Default NONE