RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-24
The DEF is 1.
Example
Setting the AC voltage measurement the reading resolution to 5 ½ :
:RESOlution:CURRent:AC 2
Syntax
:RESOlution:RESistance?
:RESOlution:RESistance {0|1|2|MIN|MAX|DEF}
Function
Sets the reading resolution for 2-wire resistance measurements.
The query returns 0, 1 or 2.
Explanation
The 2-wire resistance measurement function must be enabled
before using this command.
For the reading resolution of each parameter, please refer to
the Explanationin :RESOlution:VOLTage:DC.
The DEF is 1.
Example
Setting the reading resolution of 2-wire resistance measurement to
5 ½ :
:RESOlution:RESistance 1
Syntax
:RESOlution:FRESistance?
:RESOlution:FRESistance {0|1|2|MIN|MAX|DEF}
Function
Sets the reading resolution for 4-wire resistance measurements.
The query returns 0, 1 or 2.
Explanation
The 4-wire resistance measurement function must be enabled
before using this command.
For the reading resolution of each parameter, please refer to
the Explanationin :RESOlution:VOLTage:DC.
The DEF is 1.
Example
Setting the reading resolution of 4-wire resistance measurement to
5 ½ :
:RESOlution:FRESistance 1
Syntax
:RESOlution:CAPacitance?
:RESOlution:CAPacitance {0|1|2|MIN|MAX|DEF}
Function
Sets the reading resolution for capacitance measurements.
The query returns 0, 1 or 2.