Rigol DM3064 Spezifikationen Seite 65

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 87
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 64
Chapter 2 Command System RIGOL
Programming Guide for DM3000
2-53
Explanation
<Cycles> ranges from 1 to 10000.
10. :SCAN:STOP
Syntax
:SCAN:STOP
Function
Stops the current scan task.
11. :SCAN:DATA?
Syntax
:SCAN:DATA?
<value1>,<value2>
Function
The query returns the scan data with a specified number and
quantity in the memory.
Explanation
The scan data are saved from number 1 after the start of scan.
<value1>
defines the start number of returned data.
<value2>
defines the data quantity (within 1 and 100) to be
returned.
Example
Queries the three data from number 2:
:SCAN:DATA? 2,3
The query returns:
1.36941690e-02, 1.36941690e-02, 1.36941690e-02
12. :SCAN:CARDid?
Syntax
:SCAN:CARDid?
Function
Queries the installed scan board ID number.
The query returns NONE if no boards are installed.
Seitenansicht 64
1 2 ... 60 61 62 63 64 65 66 67 68 69 70 ... 86 87

Kommentare zu diesen Handbüchern

Keine Kommentare