Seitenansicht 41
RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-30
Syntax
:SYSTem:TYPE?
Function
The query returns the instrument type by a string such as
DM3064.
Syntax
:SYSTem:SERIal?
Function
The query returns the instrument serial number by a string such as
DM3A083100011.
Syntax
:SYSTem:SCANserial?
Function
The query returns the serial number of the scan module inside the
instrument by a string. If the meter does not install any scan
module, the query returns NONE.
Syntax
:SYSTem:OPENtimes?
Function
The query returns the number of power-on such as 61.
Syntax
:SYSTem:ERRor?
Function
The query returns the error queue. If there is no error information,
the query returns: 0, "No error" (with quotation marks).
Syntax
:SYSTem:VERSion?
Function
The query returns the version number of SCPI commands: 1999.0.