RIGOL 2 Command System
2-138 MSO2000A/DS2000A Programming Guide
:MASK:CREate
Syntax
:MASK:CREate
Description
Create the mask of the pass/fail test using the horizontal adjustment parameter and the vertical
adjustment parameter currently set.
Explanation
This command is valid only when the pass/fail test function is enabled (
:MASK:ENABle) and is in
running state (
:MASK:OPERate).
For the setting of the horizontal adjustment parameter, refer to the
:MASK:X command. For the
setting of the vertical adjustment parameter, refer to the
:MASK:Y command.
:MASK:PASSed?
Syntax
:MASK:PASSed?
Description
Query the number of frames that pass the pass/fail test.
Return Format
The query returns the number of frames that pass the pass/fail test in integer.
:MASK:FAILed?
Syntax
:MASK:FAILed?
Description
Query the number of failed frames in the pass/fail test.
Return Format
The query returns the number of failed frames in the pass/fail test in integer.