RIGOL 2 Command System
2-148 MSO2000A/DS2000A Programming Guide
:MEASure:SETup:MAX
:MEASure:SETup:MID
:MEASure:SETup:MIN
Syntax
:MEASure:SETup:MAX <max>
:MEASure:SETup:MAX?
:MEASure:SETup:MID <mid>
:MEASure:SETup:MID?
:MEASure:SETup:MIN <min>
:MEASure:SETup:MIN?
Description
Set or query the threshold level upper limit, middle value or lower limit of the auto measurement of the
analog channels. The unit is %.
Parameter
Name Type Range Default
<max> Integer 7 to 95 90
<mid> Integer 6 to 94 50
<min> Integer 5 to 93 10
Explanation
The upper limit, middle value and lower limit of the threshold level are expressed in percentage of
the amplitude and are used to define the time, delay and phase parameters. Setting the threshold
will affect all the time, delay and phase parameters. You can also use the
:MEASure:SETup:TYPE
command to select the threshold measurement setting.
The middle value set must be lower than the upper limit currently set and greater than the lower limit
currently set.
Return Format
The query returns the threshold level upper limit, middle value or lower limit in integer.