RIGOL 2 Command System
2-176 MSO2000A/DS2000A Programming Guide
:MEASure:PREShoot:SMAXimum
:MEASure:PREShoot:SMINimum
:MEASure:PREShoot:SCURrent
:MEASure:PREShoot:SAVerage
:MEASure:PREShoot:SDEViation
Syntax 1
:MEASure:PREShoot:SMAXimum <chan>
:MEASure:PREShoot:SMINimum <chan>
:MEASure:PREShoot:SCURrent <chan>
:MEASure:PREShoot:SAVerage <chan>
:MEASure:PREShoot:SDEViation <chan>
Description 1
Enable the preshoot measurement function of the specified channel and the statistic function of the
measurement results.
Syntax 2
:MEASure:PREShoot:SMAXimum? <chan>
:MEASure:PREShoot:SMINimum? <chan>
:MEASure:PREShoot:SCURrent? <chan>
:MEASure:PREShoot:SAVerage? <chan>
:MEASure:PREShoot:SDEViation? <chan>
Description 2
Query the statistic maximum value, minimum value, current value, average value or standard deviation of
the preshoot measurement results when the preshoot measurement function of the specified channel and
the statistic function of the measurement results are enabled.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the statistic maximum value, minimum value, current value, average value or standard
deviation of the preshoot measurement results in scientific notation. When the measurement results are
invalid, the query returns 9.9e37.