RIGOL
Programming Guide for DM3058/DM3058E
4-7
7. CALCulate:NULL:OFFSet
Command Format:
CALCulate:NULL:OFFSet {<value>|MINimum|MAXimum}
CALCulate:NULL:OFFSet? [MINimum|MAXimum]
Function:
The command commands set and query the offset of NULL operation.
Parameter:
Name Type Range
<value>
Consecutive
Integer
±120% of MAX range
Explanations:
The command is valid only when the NULL function (CALCulate:FUNCtion NULL)
is enabled under calculation is active (CALCulate:STATe ON).
CALCulate:NULL:OFFSet {<value>|MINimum|MAXimum} - Sets the desired
NULL offset.
CALCulate:NULL:OFFSet? [MINimum|MAXimum] - Queries the current NULL
offset.
MIN denotes the minimum range (-120% of MAX range) of current measurement.
MAX denotes the maximum range (120% of MAX range) of current measurement.
Return Format:
The query returns the current NULL offset.
8. CALCulate:AVERage:AVERage?
Command Format:
CALCulate:AVERage:AVERage?
Function:
The command queries the average of all data after calculation.
Explanations:
The command is valid only when the calculation function (CALCulate:FUNCtion
AVERage) is enabled under calculation is active (CALCulate:STATe ON).