Rigol DM3058 Bedienungsanleitung Seite 159

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 192
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 158
RIGOL
Programming Guide for DM3058/DM3058E
5-10
The command enables the multimeter exit Touch Hold mode and back to its
normal state.
6. HOLDTHRESH
Command Format:
HOLDTHRESH <threshold>
HOLDTHRESH?
Function:
The commands set and query the threshold for HOLD measurement.
Explanations:
HOLDTHRESH <threshold> - Sets the HOLD threshold to <threshold>,
<threshold> should be 1, 2, 3 or 4 (which are 0.01%, 0.1%, 1%, 10%,
respectively), or else errors may occur.
HOLDTHRESH? - Queries the threshold of HOLD measurement.
Return Format:
The query returns the HOLD threshold 1, 2, 3 or 4.
7. MAX
Command Format:
MAX
Function:
The command enables the multimeter enter MAX operation with current reading
as maximum.
Explanations:
If the multimeter has been in MAX mode, the maximum value will be shown on
the screen. Note the auto range function is disabled in this mode.