Programming Guide for DM3058/DM3058E
4-34
Function:
The commands set and query the range of AC current measurement.
Parameter:
Name Type Range Default
<range> Discrete { 20 mA|200 mA|2 A|10 A } 200 mA
Explanations:
[SENSe:]CURRent:AC:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of AC current measurement from {20 mA|200 mA|2 A|10 A }; thereinto,
MIN = 20 mA, MAX = 10 A and the unit is A. If a 200 mA range is required, enter
0.2 as its parameter.
[SENSe:]CURRent:AC:RANGe? [MINimum|MAXimum] Queries the current
range of AC current measurement.
Return Format:
The query returns the range of AC current measurement.
6. [SENSe:]RESistance:RANGe
Command Format:
[SENSe:]RESistance:RANGe {<range>|MINimum|MAXimum}
[SENSe:]RESistance:RANGe? [MINimum|MAXimum]
Function:
The commands set and query the range of 2-wire resistance measurement.
Parameter:
Name Type Range Default
<range> Discrete {200 Ω|2 kΩ|20 kΩ|200 kΩ
|2 MΩ|10 MΩ|100 MΩ}
200 kΩ
Explanations:
[SENSe:]RESistance:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of 2-wire resistance measurement from {200 Ω|2 kΩ|20 kΩ|200 kΩ|2
MΩ|10 MΩ|100 MΩ}; thereinto, MIN = 200 Ω, MAX = 100 and the unit is Ω. If
a 2 kΩ range is required, enter 2000 as its parameter.