Rigol DM3058 Bedienungsanleitung Seite 47

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 192
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 46
RIGOL
Programming Guide for DM3058/DM3058E
3-22
Explanations:
The available measurements to main-display could only be DCV, DCI, ACV, ACI
after this command.
7. :FUNCtion2:CURRent:AC
Command Format:
:FUNCtion2:CURRent:AC
Function:
The command sets the vice-display to AC current measurement.
Explanations:
The available measurements to main-display could only be DCV, DCI, ACV, ACI,
FREQUENCY, PERIOD after this command.
8. :FUNCtion2:FREQuency
Command Format:
:FUNCtion2:FREQuency
Function:
The command sets the vice-display to frequency measurement.
Explanations:
The available measurements to main-display could only be ACV, FREQUENCY,
PERIOD after this command.
9. :FUNCtion2:PERiod
Command Format:
:FUNCtion2:PERiod
Function:
The command sets the vice-display to period measurement.