Programming Guide for DM3058/DM3058E
4-37
Parameter:
Name Type Range Default
<range> Discrete {200 mV|2 V|20 V
|200 V|750 V}
20 V
Explanations:
[SENSe:]PERiod:VOLTage:RANGe {<range>|MINimum|MAXimum} - Sets the
desired voltage range under period measurement from {200 mV|2 V|20 V|200
V|750 V}; thereinto, MIN = 200 mV, MAX = 750 V and the unit is V. If a 200 mV
range is required, enter 0.2 as its parameter.
[SENSe:]PERiod:VOLTage:RANGe? [MINimum|MAXimum] Queries the current
voltage range under period measurement.
Return Format:
The query returns the voltage range under period measurement.
10. [SENSe:]VOLTage:DC:RANGe:AUTO
Command Format:
[SENSe:]VOLTage:DC:RANGe:AUTO {ON|OFF}
[SENSe:]VOLTage:DC:RANGe:AUTO?
Function:
The commands set and query the auto range state of DC voltage measurement.
Explanations:
[SENSe:]VOLTage:DC:RANGe:AUTO {ON|OFF} - Sets the desired auto range
state. If choose ON, the instrument will select a measurement range according to
the signal inputted.
[SENSe:]VOLTage:DC:RANGe:AUTO? Queries the current auto range state of
DC voltage measurement.
Return Format:
The query returns auto range state.