Function:
The command resets all the period measuring parameters and trigger parameters
as defaults, and sets period measuring function.
Parameters:
Name
Type
Range of Vaules
Default
<range>
Consecutive
Integer
1 μs~50 ms
50 ms
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 1 μs~50 ms (unit is s). MAX = 0.05, MIN = 0.000001, DEF =
0.05.
<resolution> (unit is Hz); in default setting, the rate is medium.
10. CONFigure:RESistance
Command Format:
CONFigure:RESistance
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the 2-wire resistance measuring parameters and trigger
parameters as defaults, and sets 2-wire resistance measuring function.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 Ω | 2 kΩ | 20 kΩ | 200 kΩ
| 2 MΩ | 10 MΩ | 100 MΩ}
2
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 200 Ω (MIN), 2 , 20 , 200 , 2 , 10 , 100 (MAX),
AUTO (autorange) (unit is Ω). MAX = 100 000 000, MIN = 200.
<resolution> (unit is Ω); in default setting, the rate is medium.