Rigol DM3058 Spezifikationen Seite 180

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 264
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 179
RIGOL
Programming Guide for DM3058
The query returns the enable state of DCV NULL function.
27. [SENSe:]VOLTage[:DC]:NULL:VALue
Command Format:
[SENSe:]VOLTage[:DC]:NULL:VALue{<value>|MIN|MAX}
[SENSe:]VOLTage[:DC]:NULL:VALue?[{MIN|MAX}]
Function:
The command sets or queries the NULL value of DCV measurement.
Parameters:
Name
Type
Range of Values
Default
<value>
Discrete
± 120% max range
0
Explanations:
[SENSe:]VOLTage[:DC]:NULL:VALue{<value>|MIN|MAX}: Set the NULL value
of DCV measurement. MIN is -120% max range of current function, MAX is 120%
max range of current function, DEF = 0.
[SENSe:]VOLTage[:DC]:NULL:VALue?[{MIN|MAX}]: Query the NULL value of
DCV measurement.
Before sending the command, do run the command [SENSe:]VOLTage[:DC]:
NULL[:STATe]ON to enable NULL function.
Return Format:
The query returns the NULL value of DCV measurement.
28. [SENSe:]VOLTage[:DC]:PEAK:STATe
Command Format:
[SENSe:]VOLTage[:DC]:PEAK:STATe{ON|OFF}
[SENSe:]VOLTage[:DC]:PEAK:STATe?
Function:
The command sets or queries the enable state of DCV peak-to-peak measuring
function.
Seitenansicht 179
1 2 ... 175 176 177 178 179 180 181 182 183 184 185 ... 263 264

Kommentare zu diesen Handbüchern

Keine Kommentare