RIGOL
Programming Guide for DM3058
Explanations:
[SENSe:]CURRent[:DC]:NULL[:STATe]{ON|OFF}: Enable or disable DCI NULL
function.
[SENSe:]CURRent[:DC]:NULL[:STATe]?: Query the enable state of DCI NULL
function.
The command [SENSe:]CURRent[:DC]:NULL:VALue can be used to set NULL
value.
The commands CONFigure:CURRent:DC and MEASure:CURRent:DC? can
disable NULL function (OFF).
Return Format:
The query returns the enable state of DCI NULL function.
11. [SENSe:]CURRent[:DC]:NULL:VALue
Command Format:
[SENSe:]CURRent[:DC]:NULL:VALue{<value>|MIN|MAX}
[SENSe:]CURRent[:DC]:NULL:VALue?[{MIN|MAX}]
Function:
The command sets or queries the NULL value of DCI measurement.
Parameters:
Name
Type
Range of Values
Default
<value>
Discrete
± 120% max range
0
Explanations:
[SENSe:]CURRent[:DC]:NULL:VALue {<value>|MIN|MAX}: Set the NULL
valueof DCI measurement. MIN is -120% max range of current function, MAX is
120% max range of current function, DEF = 0.
[SENSe:]CURRent[:DC]:NULL:VALue?[{MIN|MAX}]: Query the NULL value of
DCI measurement.
Before sending the command, do run the command [SENSe:]CURRent[:DC]:
NULL[:STATe]ON to enable NULL function.
Return Format:
The query returns the NULL value of DCI measurement.