5-59
Return Format:
The query returns the NULL value of ACV measurement.
19. [SENSe:]VOLTage:AC:PEAK:STATe
Command Format:
[SENSe:]VOLTage:AC:PEAK:STATe{ON|OFF}
[SENSe:]VOLTage:AC:PEAK:STATe?
Function:
The command sets or queries the enable state of ACV peak-to-peak measuring
function.
Explanations:
[SENSe:]VOLTage:AC:PEAK:STATe {ON|OFF}: Enable or disable ACV
peak-to-peak measuring function.
[SENSe:]VOLTage:AC:PEAK:STATe?: Query the enable state of ACV peak-to-peak
measuring function.
The command ―[SENSe:]VOLTage:AC:NULL:VALue‖ can be used to set ACV
peak-to-peak value.
Return Format:
The query returns the enable state of ACV peak-to-peak measuring function.
20. [SENSe:]VOLTage:AC:RANGe:AUTO
Command Format:
[SENSe:]VOLTage:AC:RANGe:AUTO{ON|OFF}
[SENSe:]VOLTage:AC:RANGe:AUTO?
Function:
The command sets or queries the enable state of ACV autorange function.
Explanations:
[SENSe:]VOLTage:AC:RANGe:AUTO {ON|OFF}: Enable or disable ACV
autorange function. If ACV autorange function is ON, the measuring range will be
selected according to input signal.