RIGOL
Programming Guide for DM3058
5-61
Function:
The command sets or queries the interval time of DCV measurement.
Parameters:
Name
Type
Range of Values
Default
<seconds>
Consecutive
Real Number
400 ms~2 s
400 ms
Explanations:
[SENSe:]VOLTage[:DC]:APERture{<seconds>|MIN|MAX|DEF}: Set the interval
time of DCV measurement. Unit is s, so 400 ms should be input in the form of
0.4.
[SENSe:]VOLTage[:DC]:APERture? [{MIN|MAX}]: Query the interval time of
DCV measurement.
Return Format:
The query returns current interval time of DCV measurement.
23. [SENSe:]VOLTage[:DC]:APERture:ENABled
Command Format:
[SENSe:]VOLTage[:DC]:APERture:ENABled{ON|OFF}
[SENSe:]VOLTage[:DC]:APERture:ENABled?
Function:
The command sets or queries the enable state of DCV interval time measuring
function.
Explanations:
[SENSe:]VOLTage[:DC]:APERture:ENABled{ON|OFF}: Enable oe disable DCV
interval time measuring function.
[SENSe:]VOLTage[:DC]:APERture:ENABled?: Query the enable state of DCV
interval time measuring function.
The command ―CONFigure:VOLTage:DC‖, ―MEASure:VOLTage:DC?‖, ―[SENSe:]
VOLTage[:DC]:NPLC‖ and ―[SENSe:]VOLTage[:DC]:RESolution‖ can be used to
disable interval time measuring function (OFF).