Programming Guide for DM3058
14. :calculate:DBM:state
Command Format:
:calculate:DBM:state?
:calculate:DBM:state{ON|OFF}
Function:
The command queries or sets the enable state of dBm operation.
Explanations:
:calculate:DBM:state?: Query the enable state of dBm operation.
:calculate:DBM:state{ON|OFF}: Enable or disable dBm operation.
Return Format:
The query returns the enable state of dBm operation: ON or OFF that denotes
turning dBm operation on or off.
Example:
Disable Stat. Operation:
:calculate:DB:stateOFF
15. :calculate:limit?
Command Format:
:calculate:limit?
Function:
The command queries the result of P/F calculation.
Explanation:
When the measuring value is between min and max, the command returns pass,
or else fail.
Return Format:
The query returns whether P/F calculation have passed: PASS or FAIL.